Analysis of water hardness (total and residual hardness) with good/ non good water signal.
Colorimetric analyzer, with microprocessor for water hardness control with the following alarm set points (the set point is established by the indicator type):
Technical features
Power supply:
230V 50-60Hz (85-264 V, 47-63 Hz)
24V AC-DC
Current consumption: 25 VA
Degree of protection with support: IP 43
Degree of protection with mounting box: IP 54
Amount of reagent: 500 ml
Reagent consumption: 5000 analysis with 1 bottle (500ml)
Inputs: 1 input, maximum load 18V 12mA (STOP or START analysis, water meter)
Relay outputs: 3 outputs, maximum load 250V 4A (maximum value, alarm, analysis in force)
SD card (results, errors)
Analysis interval: water meter or flow switch
Washing time: 5, 10, 20 or 30 minutes
Quality of the water to be analyzed
Temperature: 5° - 40° C
Optically: transparent, without color, without suspended substances, without air bubbles and CO2
Chemically: pH 4 -10.5, iron <3ppm, copper <0.2ppm, aluminum <0.1ppm, manganese <0.2ppm acid capacity K S4.3 <5mmol / l
Water pressure: 0,2-6 bar
In / out water connection: 6 mm flexible tube outer diameter